Usability Engineering File Template (IEC 62366-1)

IEC 62366-1 usability file: use specification, known use problems, hazard-related use scenarios, formative and summative evaluation, root causes

ISO & KonformitätVollständigIEC 62366-1:2015+A1:2020usabilityengineeringfile

Anleitung: Use it when you run usability engineering for a medical device and need one file that links use errors to the risk file. Fill the use specification and the hazard-related use scenarios first, then plan the summative evaluation before you see results. The main pitfall is reporting a pass rate without a root cause analysis of every use error and close call. Honest current-state scores matter more than target ambitions.

Vorschau

Usability Engineering File: [Device name and model]

The usability engineering file for one medical device, structured on IEC 62366-1:2015 with Amendment 1:2020 (Medical devices, Part 1: Application of usability engineering to medical devices, consolidated edition 1.1, June 2020). It records the use specification, the user interface characteristics related to safety, known use problems, hazard-related use scenarios, the user interface specification, the evaluation plan, formative evaluations and the summative evaluation. The process focuses on use errors that can lead to harm in normal use; it does not try to prevent deliberate abnormal use. Replace every [bracketed] value and every row marked Example. This is a documentation format, not regulatory or medical advice: check the current official texts, the expectations of your target markets and a qualified human factors and regulatory professional.

Document control

FieldValue
Document ID[UEF-NNN]
Device[Name, model, variants]
User interface version[Hardware, software and labelling versions evaluated]
Linked records[Risk management file ID, design and development file ID]
Version[0.1]
Status[Draft / In review / Approved]
Owner[@usability-lead]
Last updatedYYYY-MM-DD
Classification[Internal / Confidential]

Approvals

RoleNameDecisionDate
Usability engineering lead[@name][Approved]YYYY-MM-DD
Risk management lead[@name][Approved]YYYY-MM-DD
Clinical expert[@name][Approved]YYYY-MM-DD

Revision history

VersionDateAuthorChange summary
0.1YYYY-MM-DD[@name]Use specification and known use problems
0.6YYYY-MM-DD[@name]Example: Formative evaluations FE-01 to FE-03 added
1.0YYYY-MM-DD[@name]Example: Summative evaluation USAB-07 report and conclusion

Contents

  1. Use specification (cl. 5.1)
  2. User interface characteristics related to safety and potential use errors (cl. 5.2)
  3. Known or foreseeable hazards and hazardous situations (cl. 5.3)
  4. Hazard-related use scenarios (cl. 5.4)
  5. Selection for summative evaluation (cl. 5.5)
  6. User interface specification (cl. 5.6)
  7. User interface evaluation plan (cl. 5.7)
  8. Formative evaluations (cl. 5.8)
  9. Summative evaluation (cl. 5.9)
  10. Process, tailoring and user interface of unknown provenance (cl. 4, Annex C)
  11. Traceability matrix
  12. Glossary and references

1. Use specification (cl. 5.1)

ItemValue
Intended medical indicationExample: Continuous or intermittent intravenous infusion of fluids and medicines
Patient populationExample: Adults and children above 10 kg in hospital
Part of the body or tissueExample: Venous access through a standard administration set
Operating principleExample: Volumetric peristaltic pump with drug library and dose limits
Use environmentExample: Intensive care units, general wards, in-hospital transport; noise up to ward alarm levels, variable lighting, gloves worn
User groupDescriptionTraining assumedMain tasks
UG-01 (Example)Intensive care nursesDevice training at induction, frequent useProgram, titrate, respond to alarms
UG-02 (Example)General ward nursesDevice training at induction, occasional useProgram standard infusions, respond to alarms
UG-03 (Example)Biomedical techniciansService trainingSet-up, drug library upload, preventive maintenance

2. User interface characteristics related to safety and potential use errors (cl. 5.2)

IDUser interface characteristicPotential use errorError type
UIC-01 (Example)Numeric keypad with decimal pointEnters 10 times the intended rateAction slip
UIC-02 (Example)Unit selector (mL/h, mg/kg/min)Programs in the wrong unitCognition, mistake
UIC-03 (Example)Administration set loading doorLoads the set without closing the clamp, free flowAction, omission
UIC-04 (Example)Occlusion alarm tone and displayAlarm not noticed in a noisy wardPerception
UIC-05 (Example)Drug library search listSelects a drug with a similar namePerception, confusion

3. Known or foreseeable hazards and hazardous situations (cl. 5.3)

IDSourceKnown use problemHazard (RMF)
KUP-01 (Example)Published infusion error studiesDecimal and tenfold programming errorsHZ-04
KUP-02 (Example)Field safety notices for similar pumpsFree flow during set loadingHZ-07
KUP-03 (Example)Own complaints, previous modelOcclusion alarm missed in noisy wardHZ-06
KUP-04 (Example)Medication safety reportsLook-alike drug names in librariesHZ-04

4. Hazard-related use scenarios (cl. 5.4)

IDTask sequenceUse errorHazardous situationHarmSeverity (RMF)
HRUS-01 (Example)Select drug, enter rate, confirm, startRate entered 10 times too highOver-dose deliveredSerious injury4
HRUS-02 (Example)Open door, load set, close door, open clampClamp opened before the set is seatedUncontrolled free flowSerious injury4
HRUS-03 (Example)Hear alarm, identify occlusion, clear it, restartAlarm not noticed for 10 minTherapy interruptedSerious injury for critical drugs4
HRUS-04 (Example)Search library, select drugLook-alike drug selectedWrong limits appliedSerious injury4
HRUS-05 (Example)Upload drug libraryOutdated library file uploadedWrong limits appliedSerious injury4

5. Selection for summative evaluation (cl. 5.5)

ScenarioIncluded?Reason
HRUS-01 (Example)YesSerious harm, primary programming task
HRUS-02 (Example)YesSerious harm, known field problem
HRUS-03 (Example)YesSerious harm, own complaint history
HRUS-04 (Example)YesSerious harm
HRUS-05 (Example)NoLibrary upload software checks version and signature (SRS-075); error verified impossible by test ST-260, argued in the risk file

6. User interface specification (cl. 5.6)

IDRequirementAddressesVerification
UIS-01 (Example)Rates above the soft limit need a second confirmation screen showing the value in large digitsHRUS-01USAB-07 task 3, ST-230
UIS-02 (Example)The pump door cannot close unless the anti-free-flow clip is engagedHRUS-02Design verification VER-140, USAB-07 task 1
UIS-03 (Example)High-priority alarm volume adjustable but never below a set minimum, with flashing light visible from 4 mHRUS-03Alarm test, USAB-07 task 5
UIS-04 (Example)Drug names in the library shown with tall-man lettering for look-alike pairsHRUS-04USAB-07 task 2
UIS-05 (Example)Quick reference card covers loading and alarm responseHRUS-02, HRUS-03Labelling review

7. User interface evaluation plan (cl. 5.7)

7.1 Formative evaluation plan

IDPurposeMethodStageParticipants
FE-01 (Example)Explore programming flowsCognitive walkthrough with paper prototypesConcept6 nurses
FE-02 (Example)Compare two confirmation screen designsSimulated use on a prototypeDetailed design8 nurses
FE-03 (Example)Check alarm perception in ward noiseSimulated ward, recorded background noisePre-verification10 nurses

7.2 Summative evaluation plan

ItemPlan
Study IDUSAB-07 (Example)
ScenariosHRUS-01 to HRUS-04
ParticipantsExample: 15 per user group: UG-01 15, UG-02 15, UG-03 15, total 45 (UG-03 performs set-up and HRUS-02 tasks only)
TrainingExample: Same training as real users receive, with a gap of at least 1 day before the session to allow for training decay
EnvironmentExample: Simulated ward with background noise and realistic lighting
DeviceProduction-equivalent units, released software and final labelling
Data collectedObserved use errors, close calls, use difficulties, task completion, participant comments in an interview after the tasks
AnalysisRoot cause analysis of every use error and close call; assessment of residual risk in the risk management file
AcceptanceNo use error on a selected scenario whose residual risk is judged not acceptable after root cause analysis

Participant numbers: check the expectations of each target market. The US FDA guidance Applying Human Factors and Usability Engineering to Medical Devices (issued 3 August 2026) recommends, in general, at least 15 participants from each distinct user population for human factors validation testing, possibly more for some device types; check the current text before planning.

8. Formative evaluations (cl. 5.8)

IDDateKey findingsDesign changeLinked requirement
FE-01 (Example)YYYY-MM-DD4 of 6 nurses looked for the unit selector after typing the rateUnit chosen before rate entryUIS-01
FE-02 (Example)YYYY-MM-DDDesign B caught 7 of 8 simulated tenfold entries, design A caught 3 of 8Design B adoptedUIS-01
FE-03 (Example)YYYY-MM-DD3 of 10 nurses did not notice the alarm within 2 min at minimum volumeMinimum volume raised, light addedUIS-03

9. Summative evaluation (cl. 5.9)

9.1 Results by scenario

ScenarioParticipantsUse errorsClose callsUse difficulties
HRUS-01 (Example)30 (UG-01, UG-02)123
HRUS-02 (Example)45 (all groups)014
HRUS-03 (Example)30 (UG-01, UG-02)002
HRUS-04 (Example)30 (UG-01, UG-02)111

9.2 Root cause analysis

ObservationScenarioParticipantRoot causeConsequence for riskAction
OBS-01 (Example): tenfold rate confirmed on the confirmation screenHRUS-01P-12 (UG-02)Participant read the drug name, not the rate, on the confirmation screenThe rate was above the hard limit, so the pump refused it (SRS-052); for a drug with wider limits the error could reach the patient, residual risk reviewed in RMF HZ-04Rate moved to top of screen, retested in USAB-07b with 15 UG-02 participants: 0 use errors
OBS-02 (Example): look-alike drug selectedHRUS-04P-27 (UG-01)Search list sorted by recent use placed the wrong drug firstLimits of the wrong drug were in a similar range; harm judged possibleRecent-use sorting removed, retested in USAB-07b: 0 use errors
OBS-03 (Example): close call, clamp almost opened earlyHRUS-02P-38 (UG-03)Habit from previous pump modelAnti-free-flow clip prevented flowNo change; covered by UIS-02 design control

9.3 Conclusion

[State whether the user interface can be used safely for the intended use, which residual use-related risks remain, and where they are disclosed.] Example: After the changes verified in USAB-07b, no use error remained on any selected scenario; residual use-related risks are documented in the risk management file and disclosed in the instructions for use.

10. Process, tailoring and user interface of unknown provenance (cl. 4, Annex C)

ItemValue
Process procedure[QMS procedure ID]
Tailoring of effort (cl. 4.3)Example: Full process for the new pump user interface; reduced process for the unchanged administration set labelling, justified by 5 years of complaint data
Information for safetyExample: Warnings on free flow and alarm volume in the instructions for use, with their effectiveness evaluated in USAB-07
User interface of unknown provenance (Annex C)[None / which parts, and the evidence reviewed]
Accompanying documentation and training evaluated[Yes / No, study IDs]

11. Traceability matrix

ScenarioHazard (RMF)UI characteristicRequirementFormativeSummativeResidual risk
HRUS-01 (Example)HZ-04UIC-01, UIC-02UIS-01FE-01, FE-02USAB-07, USAB-07bAcceptable
HRUS-02 (Example)HZ-07UIC-03UIS-02, UIS-05FE-02USAB-07Acceptable
HRUS-03 (Example)HZ-06UIC-04UIS-03, UIS-05FE-03USAB-07Acceptable
HRUS-04 (Example)HZ-04UIC-05UIS-04FE-01USAB-07, USAB-07bAcceptable

12. Glossary and references

TermWorking meaning in this file
Use errorAn action or lack of action by the user that leads to a result different from what the manufacturer intended or the user expected
Close callA use error that almost happened, or was caught and corrected before it caused a result
Use difficultyA task the user completed with noticeable effort, hesitation or help
Hazard-related use scenarioA sequence of tasks in which a use error could lead to a hazardous situation
Formative and summative evaluationEvaluation during design to find problems, versus evaluation of the final user interface to show it can be used safely
ReferenceEdition or dateUse
IEC 62366-1, Medical devices, Part 1: Application of usability engineering to medical devices2015 + A1:2020 (consolidated edition 1.1, 2020-06-17)Process structure and clause numbers
IEC TR 62366-2, guidance on usability engineering for medical devices2016Methods guidance
ISO 149712019Risk management link
EU example: Regulation (EU) 2017/745, Annex I (sections 5 and 22 on use error and lay users)Check the consolidated textGeneral safety and performance requirements
US example: FDA, Applying Human Factors and Usability Engineering to Medical DevicesFinal, issued 3 August 2026 (first issued 3 February 2016)https://www.fda.gov/regulatory-information/search-fda-guidance-documents/applying-human-factors-and-usability-engineering-medical-devices
US example: FDA, Content of Human Factors Information in Medical Device Marketing SubmissionsFinal, May 2026https://www.fda.gov/regulatory-information/search-fda-guidance-documents/content-human-factors-information-medical-device-marketing-submissions
UK example: MHRA guidance on human factors and usability engineering for medical devices in Great BritainCheck the current MHRA pageGreat Britain expectations

Standard and licence note

Structure aligned with IEC 62366-1:2015+AMD1:2020 (consolidated edition 1.1, published 2020-06-17; stability date 2028 on the IEC webstore, checked October 2026). Clause numbers are given for orientation only. All wording, tables, IDs and examples are original; no text, table or definition is reproduced from the standard. The device, participants and results are fictional examples. This template is not regulatory or medical advice: check the current official texts, the expectations of your target markets and a qualified human factors and regulatory professional.

Official text: IEC, purchase or licensed access: https://webstore.iec.ch/en/publication/67220 (also available from national standards bodies).

Template licence: MIT.