Design & Development Record (Medical Device)
Medical device design record (DHF)
Cómo usar: The input→output→V&V→transfer chain with traceability is the whole template. Every design change re-enters the chain – auditors test the last change, not the first.
Vista previa
Design & Development Record (Medical Device)
ISO 13485:2016 cl.7.3 – the design controls chain: inputs → outputs → review → verification → validation → transfer → changes. Also feeds the Design History File (DHF) required by FDA 21 CFR 820.30.
| Field | Value |
|---|---|
| Device / project | name |
| Document ID | DHF-NNN |
| Risk file | ISO 14971 file ref |
| Stage gate | current phase |
1. Design inputs (7.3.3)
| ID | Requirement | Source | Acceptance criterion |
|---|---|---|---|
| DI-01 | User need | intended use | measurable criterion |
| DI-02 | Regulatory requirement | standard/reg | clause |
| DI-03 | Safety requirement | risk file | risk control |
2. Design outputs (7.3.4)
- Specifications, drawings, manufacturing instructions, labeling
- Each output traced to input – traceability matrix required
| Input | Output doc | Verified by |
|---|---|---|
| DI-01 | SPEC-NNN | test method |
3. Design reviews (7.3.5)
| Review | Date | Attendees | Inputs covered | Issues | Decision |
|---|---|---|---|---|---|
| DR-1 | proceed/hold |
4. Verification (7.3.6) – did we build it right?
| Output | Test/method | Acceptance | Result | Pass? |
|---|---|---|---|---|
| SPEC | protocol ref | criterion |
5. Validation (7.3.7) – did we build the right thing?
- Clinical evaluation / usability / performance in intended use
| Validation | Method | Sample | Result |
|---|
6. Design transfer (7.3.8)
- Outputs transferred to production: procedures, specs, training
- Production verification records
7. Design changes (7.3.9)
| Change | Rationale | Re-verification/validation needed | Approved |
|---|---|---|---|
Linkages
- Risk management file (ISO 14971) updated with each design change
- CAPA for any verification/validation failure