Method Validation & Measurement Uncertainty Record

Lab method validation + MU budget (KaTeX)

사용법: Standard methods need verification only; lab-modified methods need full validation. The MU budget's expanded U (k=2) is what goes on the report.

미리보기

Method Validation & Measurement Uncertainty Record

ISO/IEC 17025:2017 (lab competence) – cl.7.2 method selection/ verification/validation + cl.7.6 measurement uncertainty (MU). Required before a lab reports results under accreditation.

FieldValue
Methodname/SOP ref
Validation IDVAL-YYYY-NNN
Analyst@name
Approved bytechnical manager

1. Method identification

  • Standard method (verified) or non-standard/lab-developed (validated)
  • Scope: analytes, matrices, measurement range

2. Performance characteristics validated

CharacteristicExperimentAcceptance criterionResultPass?
Selectivityinterference checkno interference >N%
LinearityN-point calibrationR2 ≥ 0.99N
LOD/LOQrepeated blanksLOD ≤ N
Precision (repeatability)n≥N replicates%RSD ≤ N
Reproducibilityanalyst/day variation%RSD ≤ N
Trueness/biasCRM or spike recoveryN-N% recovery
Robustnessparameter variationstable within N%

3. Measurement uncertainty budget (cl.7.6)

U=k⋅uc,uc=∑iui2U = k \cdot u_c, \quad u_c = \sqrt{\sum_i u_i^2}

Component uiu_iTypeValueDistributionContribution
CalibrationBnormal
RepeatabilityAnormal
Purity of standardBrectangular
Combined ucu_c
Expanded UU (k=2, 95%)

4. Quality control ongoing

  • Control charts (Shewhart) per batch, acceptance rules
  • PT/ILC participation – proficiency testing records

5. Reporting statement

Result reported as: value ± U (k=2) per scope of accreditation.