Laboratory SOP Template (GLP-style)

Lab SOP with GLP-style document control: safety, reagents, steps with critical control points, QC, calculations, training

과학 연구전체OECD GLP Principles (document control conventions)labsopmath

사용법: Use to write or revise a controlled laboratory procedure. Complete the risk assessment and define critical control points with acceptance limits before the steps. Main pitfall: steps too vague for a trained newcomer to follow, and QC criteria without an action on failure.

미리보기

Standard Operating Procedure: [Procedure title]

Laboratory standard operating procedure (SOP) with document control in the style expected under Good Laboratory Practice (GLP) and ISO/IEC 17025-type quality systems: purpose, scope, responsibilities, safety, materials, procedure with critical control points, quality control, calculations, records, troubleshooting, training and revision history. This is a document format, not safety or regulatory advice; your local risk assessment and quality system take precedence. Replace every [bracketed] value and every row marked Example.

Document control

FieldValue
SOP number[SOP-LAB-0042]
Title[Procedure title]
Version[1.0]
Status[Draft / In review / Approved / Effective / Retired]
Effective dateYYYY-MM-DD
Review dueYYYY-MM-DD [typically 2 years after effective date]
Owner (author)[@name, role]
Department[Laboratory or unit]
Supersedes[SOP-LAB-0042 v0.9 or "none"]
Classification[Internal]
Copy statusControlled copy; uncontrolled when printed

Approvals

RoleNameSignatureDate
Author[@name][Signature]YYYY-MM-DD
Technical reviewer[@name][Signature]YYYY-MM-DD
Laboratory manager[@name][Signature]YYYY-MM-DD
Quality assurance[@name][Signature]YYYY-MM-DD
Health and safety officer[@name][Signature]YYYY-MM-DD

Revision history

VersionDateAuthorChange summaryRetraining required
0.9YYYY-MM-DD[@name]Draft for validation runsNot applicable
1.0YYYY-MM-DD[@name]First approved versionYes
1.1YYYY-MM-DD[@name]Example: Incubation tolerance changed from +/- 2 min to +/- 5 min after validation data reviewNo (read and acknowledge)

Contents

1. Purpose

This SOP describes how to [perform the procedure] so that results are accurate, reproducible and traceable. Example: This SOP describes how to determine total protein concentration in cell lysates using a bicinchoninic acid (BCA) microplate assay, so that downstream assays load equal amounts of protein.

2. Scope

This SOP applies to all staff performing [the procedure] in [laboratory or unit]. Use outside the validated range below requires a documented justification approved by the laboratory manager.

In scopeOut of scope
Example: Cell lysates in RIPA buffer, 20-2,000 µg/mlSamples containing more than 5 mM reducing agent
[Instrument model / location][Other instruments unless validated]
[Study types, GLP and non-GLP][Clinical diagnostic samples]

3. Definitions and abbreviations

TermMeaning in this SOP
CCPCritical control point: a step with an acceptance limit that must be checked and recorded
QCQuality control sample of known value run with each batch
BatchAll samples processed on one plate in one run
LLOQ / ULOQLower and upper limits of quantitation of the validated range
[Term][Meaning]

4. Responsibilities

RoleResponsibility
AnalystPerforms the procedure as written, records data contemporaneously, reports deviations
ReviewerChecks raw data, calculations and QC acceptance before results are released
Laboratory managerEnsures staff are trained, equipment is maintained and the SOP is current
Quality assuranceAudits compliance, approves the SOP and its revisions
Health and safety officerApproves the risk assessment and safety controls

5. Safety and hazards

All staff must read the risk assessment [RA-ID] and the safety data sheets before first use. Key hazards and controls:

HazardSourceControlPPE
Example: Skin and eye irritantBCA working reagent (alkaline)Handle in tray; avoid splashesNitrile gloves, lab coat, safety glasses
[Biological hazard][Cell lysates from BSL-2 lines][Work in biosafety cabinet][Gloves, coat]
[Physical hazard][Plate reader moving parts][Keep lid closed during run]Not applicable

Waste disposal:

WasteDisposal route
Liquid chemical waste[Labelled alkaline waste container]
Biological waste[Autoclave or chemical inactivation, then clinical waste]
Plastic consumables[Biohazard bag if contaminated, otherwise general waste]

Spills and exposure: [Follow spill procedure SOP-HS-0003; eye exposure: rinse at eyewash for 15 minutes and report the incident.]

6. Equipment and materials

6.1 Equipment

EquipmentAsset IDRequirement
Microplate reader[ID]Example: Absorbance at 562 nm, calibration within date
Incubator[ID]37 °C +/- 1 °C, verified with calibrated thermometer
Single and multichannel pipettes[IDs]Calibrated within 12 months
[Equipment][ID][Requirement]

6.2 Materials and reagents

MaterialSupplier / catalogue no.StorageNotes
[BCA reagent A and B][Supplier, cat. no.]Room temperatureRecord lot numbers
[Albumin standard, 2 mg/ml][Supplier, cat. no.]4 °CDo not freeze
[96-well flat-bottom plates][Supplier, cat. no.]Room temperatureClear polystyrene
[Material][Supplier][Storage][Notes]

7. Reagent preparation

ReagentPreparationStabilityLabel
Example: BCA working reagentMix reagent A and B at 50:1 (for one plate, 10 ml A + 200 µl B). Mix until uniformly green.24 h at room temperatureName, date, time, initials
Standard seriesDilute albumin standard in the sample buffer to 2,000, 1,500, 1,000, 750, 500, 250, 125, 25 and 0 µg/mlPrepare fresh each runName, date, initials
[Reagent][Preparation][Stability][Label]

8. Procedure

The procedure is shown as a flow, then as numbered steps.

8.1 Before starting

  1. Confirm the SOP version at the point of use is the current effective version.
  2. Check calibration labels on the plate reader, incubator and pipettes; do not proceed if any is out of date.
  3. Record reagent lot numbers and expiry dates on the run sheet [Form F-0042-1].
  4. Allow reagents and samples to reach room temperature for 30 minutes.

8.2 Procedure steps

StepActionAcceptance limitCCP
1Prepare the standard series as in section 7Visible, bubble-free volumesCCP-1
2Dilute samples 1:5 in sample buffer if expected above 2,000 µg/mlDiluted value within the standard range
3Pipette 25 µl of each standard, QC and sample into duplicate wells following the plate mapPlate map recorded before pipetting
4Add 200 µl of working reagent to each well with a multichannel pipetteWithin 5 minutes across the plate
5Cover the plate and shake for 30 s
6Incubate at 37 °C for 30 minExample: 30 min +/- 2 min, 37 °C +/- 1 °C, start and end times recordedCCP-2
7Cool to room temperature for 5 min
8Read absorbance at 562 nmReader self-check passed
9Export the raw data file to [validated location]File name per section 11
10Evaluate standard curve and QC against section 9All criteria metCCP-3

8.3 After completion

  1. Dispose of waste as in section 5.
  2. Clean the work area with [disinfectant] and record cleaning on [log].
  3. Submit the run sheet and raw data for review.

9. Quality control

Each batch includes a blank, a full standard curve and QC samples at two levels. The batch is accepted only if all criteria are met.

QC elementFrequencyAcceptance criterionAction on failure
BlankEvery batchExample: Absorbance below 0.15Reject batch; check reagent
Standard curveEvery batchR2≥0.99R^2 \ge 0.99; back-calculated standards within 15% of nominalReject batch; repeat
QC low and highEvery batch, duplicateWithin 20% of assigned valueReject batch; investigate per deviation SOP
Duplicate agreementEvery sampleCV at most 10%Repeat the sample

10. Calculations

Fit the blank-corrected standards with a [linear or four-parameter logistic] model. For a linear fit:

Acorr=A562−Ablank,C=Acorr−bm×DA_{\text{corr}} = A_{562} - A_{\text{blank}}, \qquad C = \frac{A_{\text{corr}} - b}{m} \times D

where mm is the slope, bb the intercept and DD the dilution factor.

Precision between duplicates is checked as the coefficient of variation:

CV (%)=sxˉ×100\text{CV}\,(\%) = \frac{s}{\bar{x}} \times 100

Example: With m=0.00112m = 0.00112 per µg/ml, b=0.031b = 0.031, Acorr=0.871A_{\text{corr}} = 0.871 and D=5D = 5: C=(0.871−0.031)/0.00112×5=3,750C = (0.871 - 0.031) / 0.00112 \times 5 = 3{,}750 µg/ml.

Validated calculation template: [Spreadsheet name, version, validation reference].

11. Data recording and records

RecordFormatLocationRetention
Run sheet [F-0042-1]Paper or ELN[ELN project / archive][Per retention policy, for example 10 years for GLP studies]
Raw data file[.csv / instrument format][Validated server path][Period]
Calculation file[Template][Location][Period]
Results report[PDF][LIMS][Period]

File naming: Example: SOP0042_YYYY-MM-DD_[analyst initials]_[plate number].csv. Corrections are made by a single line through the original, with the correct value, initials, date and reason. Never use correction fluid or overwrite electronic raw data.

12. Troubleshooting

ProblemLikely causeAction
High blank absorbanceExample: Contaminated or old working reagentPrepare fresh working reagent; check buffer compatibility
Poor standard curve fitPipetting error, wrong dilutionRepeat standards; check pipette calibration
Duplicates disagreeBubbles in wells, edge effectsPop bubbles with a clean tip; avoid outer wells
Sample above top standardConcentration too highDilute further and repeat
[Problem][Cause][Action]

Any failure of a CCP or QC criterion is a deviation and is recorded on [deviation form] with the impact assessment and corrective action.

Deviation handling follows these steps:

  1. Stop and make the situation safe. Do not continue a run that has failed a CCP.
  2. Record what happened, when, and which samples are affected, on the run sheet and the deviation form, the same day.
  3. Inform the reviewer and, for GLP studies, the study director, who decides on the impact on the study.
  4. Investigate the root cause and record corrective and preventive actions with owners and dates.
  5. Quality assurance reviews the deviation log at least [quarterly] and trends recurring causes.

Example: Incubator temperature found at 39 °C at the end of step 6. Batch rejected, samples re-assayed the next day from retained aliquots, incubator taken out of service and recalibrated; root cause was a faulty door seal, replaced under work order [WO-ID].

13. References

Documents referenced by this SOP:

IDDocument
REF-01[Manufacturer instructions for kit, version]
REF-02[Method validation report VAL-0042]
REF-03[Risk assessment RA-ID]
REF-04[Related SOPs: pipette calibration, deviation handling, waste disposal]
REF-05OECD Principles of Good Laboratory Practice, as applicable to the study

14. Training record

TraineeTrainerTraining methodCompetence assessed byDate signed off
[@name][@name]Example: Read SOP, observed 1 run, performed 2 supervised runs with QC passing[@name]YYYY-MM-DD
[@name][@name][Method][@name]YYYY-MM-DD

Standard and licence note

Based on common laboratory quality practice for standard operating procedures, with document control in the style expected under the OECD Principles of Good Laboratory Practice and ISO/IEC 17025-type quality systems. This template is original text written for this collection and reproduces no text from those documents; consult them and your quality system for normative requirements. This is a document format, not safety or regulatory advice. Template licence: MIT.