EU MDR Technical Documentation Template (Annex II and III)

EU MDR technical file index: classification, Annex II sections 1 to 6, GSPR checklist, V&V and clinical data, Annex III PMS plan and PSUR

ISO & Tuân thủĐầy đủRegulation (EU) 2017/745 (MDR) Annexes II and IIImdrtechnicaldocumentation
Mở trong editorTải .md

Cách dùng: Use it when you prepare or maintain the technical documentation for a device placed on the EU market under the MDR. Fill the classification (section 0) and device description (section 1) first, then point each Annex II and III section to your controlled records. The main pitfall is pasting whole reports instead of summarising them and linking the controlled document ID and version.

Xem trước

Technical Documentation: [Device name and model]

The technical documentation index and summary for one medical device (or device family under one Basic UDI-DI) placed on the EU market, structured on Annexes II and III of Regulation (EU) 2017/745 on medical devices (MDR). Annex II sets out the six parts of the technical documentation; Annex III covers the technical documentation on post-market surveillance. The file is the map a notified body or competent authority uses to find evidence: it summarises each part and points to the controlled records that hold the detail. Replace every [bracketed] value and every row marked Example. This is a documentation format, not regulatory or legal advice: the MDR is amended from time to time and a targeted revision is in the EU legislative process, so check the current consolidated text on EUR-Lex, the current MDCG guidance and a qualified regulatory professional.

Document control

FieldValue
Document ID[TD-NNN]
Device and Basic UDI-DI[Name, model, Basic UDI-DI]
Manufacturer and SRN[Legal name, address, single registration number]
Authorised representative (if outside the EU)[Name, address, SRN]
Person responsible for regulatory compliance[@name]
Risk class and rule[For example class IIb, Annex VIII rule 12]
Notified body[Name and four-digit number]
Version[0.1]
Status[Draft / In review / Approved]
Owner[@regulatory-lead]
Last updatedYYYY-MM-DD
Classification[Confidential]

Approvals

RoleNameDecisionDate
Regulatory affairs lead[@name][Approved]YYYY-MM-DD
Person responsible for regulatory compliance[@name][Approved]YYYY-MM-DD
Quality assurance[@name][Approved]YYYY-MM-DD

Revision history

VersionDateAuthorChange summary
0.1YYYY-MM-DD[@name]Structure and device description
0.8YYYY-MM-DD[@name]Example: Verification, clinical evaluation and PMS sections completed for notified body submission
1.0YYYY-MM-DD[@name]Example: Updated after notified body technical documentation assessment

Contents

  1. Regulatory strategy and classification
  2. Device description and specification (Annex II, section 1)
  3. Information to be supplied by the manufacturer (Annex II, section 2)
  4. Design and manufacturing information (Annex II, section 3)
  5. General safety and performance requirements (Annex II, section 4)
  6. Benefit-risk analysis and risk management (Annex II, section 5)
  7. Product verification and validation (Annex II, section 6)
  8. Post-market surveillance documentation (Annex III)
  9. Declaration of conformity and registrations
  10. Other markets: mapping table
  11. Glossary and references

0. Regulatory strategy and classification

ItemValue
Qualification as a medical deviceExample: Intended for the administration of medicines by infusion, a medical purpose under Article 2(1)
Classification rule(s) consideredExample: Rule 12 (active device administering medicines); rule 11 considered for the software and not applied separately because the software drives the device
Resulting classExample: IIb, because administration can be potentially hazardous given the medicines and the mode of delivery
Conformity assessment routeExample: Article 52(4) for class IIb: Annex IX chapters I (quality management system) and III (administrative provisions), with assessment of the technical documentation under Annex IX section 4 for at least one representative device per generic device group; confirm the sampling with the notified body. As a class IIb active device administering medicinal products (rule 12), the device falls under the clinical evaluation consultation procedure of Article 54(1)(b) unless an Article 54(2) exemption applies
Harmonised standards and common specifications appliedSee section 4
Transition status[New device under the MDR / legacy device under the Article 120 transitional provisions as amended; check the current dates]

1. Device description and specification (Annex II, section 1)

1.1 Description and specification

ElementContentRecord
Name and general description, intended purpose, intended usersExample: Corvane IP-200 volumetric infusion pump for continuous and intermittent intravenous infusion, used by trained nurses in hospitalsIFU [ID], use specification UEF-003 section 1
Basic UDI-DI or other identifier[Basic UDI-DI]UDI assignment record
Patient population, conditions, indications, contraindications, warningsExample: Adults and children above 10 kg; not for enteral feedingClinical evaluation plan CEP-004
Principles of operation and mode of actionExample: Peristaltic pumping mechanism, software rate control, independent safety monitorDesign description DD-010
Rationale for qualification as a deviceSee section 0
Risk class and classification ruleSee section 0
Novel featuresExample: Drug library with signed files and hard limits; none clinically novelDesign description DD-010
Accessories, other devices and products used in combinationExample: Dedicated administration sets (separate technical documentation TD-015), mains adaptorAccessory list ACC-01
Configurations and variantsExample: IP-200 (single channel), IP-200D (dual channel)Variant matrix VAR-01
Key functional elements, with diagramsExample: Pumping mechanism, main controller, safety monitor, display, batterySystem architecture SA-02
Materials in contact with the bodyExample: None directly; fluid path materials covered in TD-015Material list MAT-01
Technical specificationsExample: Rate range 0.1 to 1,200 mL/h, accuracy ±5% of set rateProduct specification PS-01

1.2 Previous and similar generations

DeviceRelationshipRelevanceRecord
Corvane IP-100 (Example)Previous generation by the same manufacturerField data used in risk and clinical evaluationPMS report PMS-IP100-YYYY
[Similar devices on the market]Comparable devices by other manufacturersState of the art, similar device dataClinical evaluation report CER-004 section [n]

2. Information to be supplied by the manufacturer (Annex II, section 2)

ItemID and versionLanguagesAnnex I chapter III check
Device label (Example)LBL-200 v3Symbols plus [languages]Checklist LC-01
Packaging label (Example)LBL-201 v2[Languages]Checklist LC-01
Instructions for use (Example)IFU-200 v4[Languages accepted in each Member State of sale]Checklist LC-02
Electronic instructions for use (if used)[ID, or not used][Check the implementing rules on electronic instructions for use]
Quick reference card (Example)QRC-200 v2[Languages]Usability file UIS-05

3. Design and manufacturing information (Annex II, section 3)

3.1 Design stages

StageOutputRecord
Design inputs (Example)User needs and design inputsDesign and development file DDF-008
Software lifecycle (Example)Software development plan, verification, releaseSDP-011
Usability engineering (Example)Use specification, formative and summative evaluationUEF-003
Design transfer (Example)Device master recordDMR-200

3.2 Manufacturing processes and validation

ProcessValidationMonitoringFinal test
PCB assembly (Example)Process validation PV-11In-line optical inspectionBoard functional test
Final assembly and flow calibration (Example)Process validation PV-14Statistical process control on flow accuracyFinal acceptance test FAT-200
Software installation (Example)Validated programming station PV-17Checksum verificationVersion check at final test

3.3 Sites, suppliers and subcontractors

Site or supplierActivityCritical?ControlCertificate
Corvane Medical Ltd, [site] (Example)Design, final assembly, releaseYesOwn QMSISO 13485 certificate [ID]
[Supplier] (Example)PCB assemblyYesSupplier audit, quality agreement[Certificate ID]
[Supplier][Activity][Yes / No][Control][Certificate]

4. General safety and performance requirements (Annex II, section 4)

Annex I sectionTopic (paraphrase)Applies?Method of conformityEvidence
1 to 9 (Example)General requirements: performance, risk management, risk reduction, use error, lifetime, transport and storageYesEN ISO 14971:2019/A11:2021, EN 62366-1 [version applied]RMF-012, UEF-003
10 (Example)Chemical, physical and biological propertiesPartlyFluid path in TD-015; housing materials assessed for user contactMAT-01
14 (Example)Construction and interaction with the environmentYesEN 60601-1 series, electromagnetic compatibility testingTest reports TR-301, TR-302
17 (Example)Electronic programmable systems and software, including IT securityYesEN 62304 [version applied], cybersecurity per IEC 81001-5-1SDP-011, security file SEC-02
18 (Example)Active devices and devices connected to them, including alarmsYesCollateral standard on alarm systems, infusion pump particular standardTR-303, TR-304
21 (Example)Devices supplying energy or substances to the patientYesFlow accuracy and over-delivery protection testingTR-305, ST-210
22 (Example)Devices for use by lay personsNoProfessional use only, stated in intended purposeIFU-200 section 1
23 (Example)Label and instructions for useYesEN ISO 15223-1 symbols, labelling checklistLC-01, LC-02

Harmonised standards: check the references published in the Official Journal under the MDR (Commission Implementing Decision (EU) 2021/1182 and its amendments) and record the exact version applied.

5. Benefit-risk analysis and risk management (Annex II, section 5)

ItemSummaryRecord
Risk management plan and fileExample: ISO 14971:2019 process, file RMF-012RMF-012
Main hazards and controlsExample: Over-dose (HZ-02, HZ-04): independent safety monitor and drug library hard limits; free flow (HZ-07): anti-free-flow clipRMF-012 sections 3 to 5
Overall residual riskExample: Acceptable when weighed against the clinical benefit of controlled infusionRMF-012 section 8
Benefit-risk conclusionExample: Benefits outweigh residual risks for the intended purposeCER-004 conclusion

6. Product verification and validation (Annex II, section 6)

6.1 Pre-clinical and clinical data

AreaTests or studiesResult summaryRecord
Electrical safety and EMC (Example)EN 60601-1 series testingPassed, no deviationsTR-301, TR-302
Performance (Example)Flow accuracy over the rate rangeWithin ±5% across the rangeTR-305
Software verification and validation (Example)System testing, traceability to requirementsAll requirements verified; known anomalies evaluatedSDP-011 sections 7 and 8
Usability (Example)Summative evaluation USAB-07 and USAB-07bNo remaining use error on selected scenariosUEF-003 section 9
Biocompatibility[Applies / not applicable with reason][Record]
Stability and shelf life (Example)Battery ageing, packaged transport testsMeets specification for the stated lifetimeTR-306
Clinical evaluation (Example)Clinical evaluation plan and report (Article 61, Annex XIV Part A)Sufficient clinical evidence for the intended purposeCEP-004, CER-004
Post-market clinical follow-up (Example)PMCF plan and evaluation report (Annex XIV Part B), or justification if not applicablePMCF survey of user sites plannedPMCF-P-004

6.2 Additional information in specific cases

Specific caseApplies?Reason or record
Device incorporating a medicinal substanceExample: NoNo medicinal substance in the device
Tissues or cells of human or animal origin, or derivativesExample: NoNone used
Substances absorbed by or dispersed in the bodyExample: NoNot applicable to an infusion pump
Substances of concern (CMR or endocrine-disrupting)Example: Not in the pump; fluid path covered in TD-015MAT-01
Sterile or defined microbiological conditionExample: No for the pump; yes for administration sets (TD-015)TD-015
Measuring function[Applies / no, with reason][Record]
Connection to other devicesExample: Yes, optional network connection to hospital systemsInteroperability verification TR-310

7. Post-market surveillance documentation (Annex III)

7.1 Post-market surveillance plan

ElementContent (Example)
Information collectedSerious incidents and field safety corrective actions, non-serious incidents and undesirable side effects, trend data, literature and databases, complaints and feedback from users, distributors and importers, public information on similar devices
Methods and analysisMonthly complaint trending, yearly literature search, statistical trend method [method]
Indicators and thresholdsOcclusion alarm complaints above 2 per 1,000 devices per year trigger risk file review; any over-infusion complaint triggers investigation
Trend reportingMethod to detect a significant increase in incidents, per Article 88
CommunicationCompetent authorities, notified body, economic operators, users
LinksCAPA procedure, traceability through UDI, PMCF plan PMCF-P-004

7.2 Periodic safety update report or PMS report

ItemValue
Report typeExample: PSUR (Article 86), as required for class IIb
Update frequencyExample: At least yearly for class IIb; check the article and current MDCG guidance
Latest report[PSUR-ID, version, date]
SubmissionExample: Made available to the notified body; check the current EUDAMED requirements

8. Declaration of conformity and registrations

ItemValueRecord
EU declaration of conformity[ID, version, date]DoC-200
Notified body certificate[Certificate number, expiry date][Record]
UDI-DI and Basic UDI-DI registration[Status][EUDAMED reference]
Summary of safety and clinical performance (Article 32)[Required / not required for this class][Record]
Retention of technical documentation[Period set by Article 10 for this device]Archive procedure

9. Other markets: mapping table

MarketFramework (as checked October 2026)Main dossierReuse from this file
European UnionRegulation (EU) 2017/745This technical documentationAll
Great BritainMedical Devices Regulations 2002 (as amended), with post-market surveillance requirements in force from 16 June 2025UKCA technical documentation, or CE marking where recognised; check the current MHRA guidance on recognition datesSections 1 to 7, with UK-specific labelling and registration
United StatesFederal Food, Drug, and Cosmetic Act; premarket notification (21 CFR Part 807) or premarket approval; QMSR (21 CFR Part 820, effective 2 February 2026)510(k), De Novo or PMA submissionVerification, software, human factors and risk evidence, rearranged to FDA formats

10. Glossary and references

TermWorking meaning in this file
Basic UDI-DIThe main identifier of a device model in the EU database, linking it to its technical documentation and certificates
General safety and performance requirementsThe requirements in Annex I that every device must meet
PMCFPost-market clinical follow-up: the continuing collection of clinical data after the device is on the market
PSURPeriodic safety update report: the summary of post-market surveillance results for class IIa, IIb and III devices
SRNSingle registration number of an economic operator in the EU database
ReferenceEdition or dateUse
Regulation (EU) 2017/745 on medical devicesAdopted 5 April 2017, amended (including by Regulation (EU) 2023/607 on transitional provisions); check the consolidated textAnnexes II and III structure, articles cited: https://eur-lex.europa.eu/eli/reg/2017/745/oj/eng
Commission proposal for a targeted revision of the MDR and IVDRPublished 16 December 2025, in the legislative process as of October 2026Watch for changes to documentation and transition rules
MDCG guidance documentsCheck the current list and revisionFor example MDCG 2019-11 (software qualification and classification) and MDCG 2022-21 (PSUR): https://health.ec.europa.eu/medical-devices-sector/new-regulations/guidance-mdcg-endorsed-documents-and-other-guidance_en
ISO 13485, ISO 14971, IEC 62304, IEC 62366-1Editions applied, see section 4Supporting processes

Standard and licence note

Based on Regulation (EU) 2017/745 (the MDR), Annex II (technical documentation) and Annex III (technical documentation on post-market surveillance), public EU law available free on EUR-Lex. The section numbering follows the Annexes; all wording is an original paraphrase and no Annex text is reproduced. The MDR has been amended several times and a targeted revision proposed by the Commission on 16 December 2025 was still in the legislative process when this template was checked in October 2026, so check the current consolidated text and the MDCG guidance before use. Paid standards named here (ISO 13485, ISO 14971, IEC 62304, IEC 62366-1, IEC 60601 series) are cited by name and number only; their official texts are available for purchase or licensed access from ISO (https://www.iso.org), IEC (https://webstore.iec.ch) and national standards bodies. The device, documents, IDs and thresholds are fictional examples. This template is not regulatory or legal advice: check the current official texts and a qualified regulatory professional.

Official text: Publications Office of the European Union, free access: https://eur-lex.europa.eu/eli/reg/2017/745/oj/eng

Template licence: MIT.